Common Problems Archives - Siglent Website - 深圳市鼎阳科技股份有限公司 //m.tapprz.com/na/category/service-support/common-problems/ Sun, 04 Jan 2026 10:04:08 +0000 en-US hourly 1 https://wordpress.org/?v=6.9.1 //m.tapprz.com/wp-content/uploads/2025/02/cropped-512X512-1-32x32.png Common Problems Archives - Siglent Website - 深圳市鼎阳科技股份有限公司 //m.tapprz.com/na/category/service-support/common-problems/ 32 32 Waveform appears but is not stable //m.tapprz.com/na/waveform-appears-but-is-not-stable/ Fri, 26 Dec 2025 09:43:25 +0000 //m.tapprz.com/?p=63180 3.1 Check that the trigger source channel is consistent with the waveform being viewed3.2 Manually adjust the LEVEL trigger knob to see whether the waveform is stable.3.3 Use AUTO to adjust the waveform automatically to determine whether the waveform is stable3.4 Adjust time base, rule out false-wave interference (such as alias displayed waveforms)

The post Waveform appears but is not stable appeared first on Siglent Website.

]]>
3.1 Check that the trigger source channel is consistent with the waveform being viewed
3.2 Manually adjust the LEVEL trigger knob to see whether the waveform is stable.
3.3 Use AUTO to adjust the waveform automatically to determine whether the waveform is stable
3.4 Adjust time base, rule out false-wave interference (such as alias displayed waveforms)

The post Waveform appears but is not stable appeared first on Siglent Website.

]]>
The oscilloscope does not turn on //m.tapprz.com/na/the-oscilloscope-does-not-turn-on/ Fri, 26 Dec 2025 09:42:55 +0000 //m.tapprz.com/?p=63179 First, check the power cord and power strip for damage. Check the power switch on the instrument to make sure it is open (SDS1000CFL series has a power switch and a power protection switch). Is the power light illuminated?Is the fan operating? If the fan works, the machine is powered on. If everything is normal […]

The post The oscilloscope does not turn on appeared first on Siglent Website.

]]>
First, check the power cord and power strip for damage. Check the power switch on the instrument to make sure it is open (SDS1000CFL series has a power switch and a power protection switch). Is the power light illuminated?
Is the fan operating? If the fan works, the machine is powered on. If everything is normal but the oscilloscope still does not turn on, please contact your nearest SIGLENT dealer.
 

The post The oscilloscope does not turn on appeared first on Siglent Website.

]]>
What does “div” mean in the specifications? //m.tapprz.com/na/what-does-div-mean-in-the-specifications/ Fri, 26 Dec 2025 09:42:18 +0000 //m.tapprz.com/?p=63178 The term “div” for most SIGLENT specifications is equal to the Vertical Scale value (Commonly shown in units of Volts/div).Take the specification for the DC Gain Accuracy of a SIGLENT SDS1052DL+For ranges ≤ 100 mV/div, the DC Gain Accuracy is given as± [3% × (|reading|+|offset|) +1% × |offset| +0.2 * div+2 mV]If:Reading = 0.5 VOffset is 0.1 VVertical scale […]

The post What does “div” mean in the specifications? appeared first on Siglent Website.

]]>
The term “div” for most SIGLENT specifications is equal to the Vertical Scale value (Commonly shown in units of Volts/div).
Take the specification for the DC Gain Accuracy of a SIGLENT SDS1052DL+
For ranges ≤ 100 mV/div, the DC Gain Accuracy is given as
± [3% × (|reading|+|offset|) +1% × |offset| +0.2 * div+2 mV]
If:
Reading = 0.5 V
Offset is 0.1 V
Vertical scale (div) = 100 mV/div = 0.1 V/div,
Then DC Gain is calculated as follows:
± [3% × (|reading|+|offset|) +1% × |offset| +0.2 * div+2 mV] =
± [3% × (|0.5 V|+|0.1 V|) +1% × |0.1 V| +0.2 div * (0.1 V/div) + 0.002 V] =
± [3% × (0.6 V) +  0.0001 V +  0.02 V + 0.002 V] = 0.018 V + 0.0001 V + 0.02 V + 0.002 V = 0.0401 V

The post What does “div” mean in the specifications? appeared first on Siglent Website.

]]>
Where do I find environmental, EMC, and other compliance information for your products? //m.tapprz.com/na/where-do-i-find-environmental-emc-and-other-compliance-information-for-your-products/ Fri, 26 Dec 2025 09:41:27 +0000 //m.tapprz.com/?p=63176 Information regarding product compliance is available on the datasheet for each product.Contact your local Siglent office for more details if you have trouble finding what you need.You can also write sales@siglent.com

The post Where do I find environmental, EMC, and other compliance information for your products? appeared first on Siglent Website.

]]>
Information regarding product compliance is available on the datasheet for each product.
Contact your local Siglent office for more details if you have trouble finding what you need.
You can also write sales@siglent.com

The post Where do I find environmental, EMC, and other compliance information for your products? appeared first on Siglent Website.

]]>
How to generate the activation code for an instrument option? //m.tapprz.com/na/how-to-generate-the-activation-code-for-an-instrument-option%ef%bc%9f/ Fri, 26 Dec 2025 09:39:56 +0000 //m.tapprz.com/?p=63151 Many SIGLENT instruments feature options that can be added by using an activation code.To generate the activation code:– Purchase the option from your favorite SIGLENT office or distributor. A full list of available options are located at the bottom of the specific product web page.– You will receive an Option Card PDF document from SIGLENT […]

The post How to generate the activation code for an instrument option? appeared first on Siglent Website.

]]>
Many SIGLENT instruments feature options that can be added by using an activation code.
To generate the activation code:
– Purchase the option from your favorite SIGLENT office or distributor. A full list of available options are located at the bottom of the specific product web page.
– You will receive an Option Card PDF document from SIGLENT that contains important information for permanent activation of the features. Keep this information handy through the next steps.
– Go to http://service.siglenteu.com/easyweb/
-Enter the information from the Option Card in the correct fields of the webpage, and press Submit:

The post How to generate the activation code for an instrument option? appeared first on Siglent Website.

]]>
What is the maximum voltage for the external trigger inputs of the SDS1X/X+, SDS2X, and SDS1202XE os //m.tapprz.com/na/what-is-the-maximum-voltage-for-the-external-trigger-inputs-of-the-sds1x-x-sds2x-and-sds1202xe-os/ Fri, 26 Dec 2025 09:34:32 +0000 //m.tapprz.com/?p=63152 The maximum safe voltage of ext trigger channel is 40 Vpp.The specification of working voltages are:Ext : ± 600 mVppExt/5: ±3 Vpp.

The post What is the maximum voltage for the external trigger inputs of the SDS1X/X+, SDS2X, and SDS1202XE os appeared first on Siglent Website.

]]>
The maximum safe voltage of ext trigger channel is 40 Vpp.
The specification of working voltages are:
Ext : ± 600 mVpp
Ext/5: ±3 Vpp.

The post What is the maximum voltage for the external trigger inputs of the SDS1X/X+, SDS2X, and SDS1202XE os appeared first on Siglent Website.

]]>
Measuring small signal waveform is disturbed too much. //m.tapprz.com/na/measuring-small-signal-waveform-is-disturbed-too-much/ Wed, 15 Jan 2025 01:22:39 +0000 //m.tapprz.com/measuring-small-signal-waveform-is-disturbed-too-much/ 1. The SDS1000 series of SIGLENT can detect millivolt signals at the minimum, and automatically turn on the bandwidth limiting function at the lowest 2mV/div gear to effectively filter out high-frequency clutter. 2. Connect the clamp to the reference ground of the signal, which can effectively reduce the waveform display distortion caused by the potential […]

The post Measuring small signal waveform is disturbed too much. appeared first on Siglent Website.

]]>
1. The SDS1000 series of SIGLENT can detect millivolt signals at the minimum, and automatically turn on the bandwidth limiting function at the lowest 2mV/div gear to effectively filter out high-frequency clutter.
2. Connect the clamp to the reference ground of the signal, which can effectively reduce the waveform display distortion caused by the potential difference between the oscilloscope and the measured signal due to the reference ground.
3. The SDS1000 series of SIGLENT can detect millivolt signals at the minimum, and automatically turn on the bandwidth limiting function at the lowest 2mV/div gear to effectively filter out high-frequency clutter.
.

The post Measuring small signal waveform is disturbed too much. appeared first on Siglent Website.

]]>
Will a bandwidth upgrade affect oscilloscope user experience? //m.tapprz.com/na/will-a-bandwidth-upgrade-affect-oscilloscope-user-experience/ Fri, 20 Sep 2024 09:26:40 +0000 //m.tapprz.com/will-a-bandwidth-upgrade-affect-oscilloscope-user-experience/ When customers purchase oscilloscopes such as SDS1104X-C and SDS1122E+, the sellers often offers bandwidth upgrade services. Some customers hesitate due to concerns about potential impacts on operational smoothness or warranty validity. In fact, upgrades performed through authorized channels will neither compromise the oscilloscopes’ performance nor affect warranty coverage… Click to view details.

The post Will a bandwidth upgrade affect oscilloscope user experience? appeared first on Siglent Website.

]]>
When customers purchase oscilloscopes such as SDS1104X-C and SDS1122E+, the sellers often offers bandwidth upgrade services. Some customers hesitate due to concerns about potential impacts on operational smoothness or warranty validity. In fact, upgrades performed through authorized channels will neither compromise the oscilloscopes’ performance nor affect warranty coverage…

Click to view details.

The post Will a bandwidth upgrade affect oscilloscope user experience? appeared first on Siglent Website.

]]>